Chapter 6—Alarms and Troubleshooting
161
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
E.99 (Arterial or SN
Blood Pump Message)
Transmit error during Flash update
while in Service Mode
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
EEPROM already used,
Power Off, Replace
EEPROM
Advisory message when uploading
hardware key option
Advisory
Put in a new hardware key or calibration EEPROM in IC 20 and
power up.
EEPROM Missing or
Reading Error
During startup, the machine cannot
properly read the EEPROM memory
chip
Opening
Screen
Message
Turn the machine off and try to power up again. If the message
repeats, take the machine out of service and alert a qualified service
technician.
Emptying
The blue dialysate line connector is
on the shunt with door closed, the red
dialysate line connector remains on
the dialyzer in order to drain the
dialysate compartment.
Warning
If this message occurs when the dialyzer is not being emptied, take
the machine out of service and alert a qualified service technician.
Emptying Stopped
When air is sensed, emptying will
stop.
Warning
Connect the red dialyzer return line to the shunt. If the warning is
repeated, take the machine out of service and alert a qualified service
technician.
Emptying too long
The dialyzer empty program has
exceeded its maximum limit.
Alarm
If blood is not sensed, return the dialyzer supply and return lines to
the shunt and close the shunt door to terminate the program. If the
machine was in dialysis (blood sensed), turn machine power off and
back on to clear the program.
Error Reading Flash
Electronic Self Test
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Fail * 9 Volt Battery
9V Power Failure Battery test has
failed.
Test
Message
Replace Battery
Fail *Actuator Arterial
High
Fail *Actuator Arterial
Low
Arterial Pressure test has failed.
Test
Message
Rerun test, if failure repeats, remove from service and alert a
qualified service technician.
Fail * Actuator
Conductivity High
Fail * Actuator
Conductivity Low
Conductivity test has failed.
Test
Message
Rerun test, if failure repeats, remove from service and alert a
qualified service technician.