Chapter 6—Alarms and Troubleshooting
183
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
UF Pump Alarm
UF pump is not connected or is not
pulsing properly.
Alarm
Press
RESET
to reset the alarm. If unable to clear the alarm, return
the blood to the patient if alarm occurs during treatment. Take
machine out of service and alert a qualified service technician.
UF Rate Error
A calculation error has been detected
Alarm
Reset the UF parameters
UF Removed cleared
This temporary message is displayed
when the Tx clock is turned on the
first time after the
New TX
key was
pressed. The UF removed has been
set to 0.
Advisory
No action necessary.
UF Removed not cleared
This temporary message is displayed
when the Tx clock is turned on other
than the first time after the
New TX
key was pressed. The UF removed
has been not been set to 0.
Advisory
No action necessary.
Upper Dia. has been set
to Min [Max]
The operator has attempted to set the
upper diastolic pressure limit higher or
lower than allowed. The machine has
set the limit to the highest or lowest
value allowed.
Dialog
Message
Verify that the limit setting is acceptable
Upper Pulse has been set
to Min [Max]
The operator has attempted to set the
upper pulse rate limit higher or lower
than allowed.
Dialog
Message
The machine has set the limit to the highest or lowest value allowed.
Verify that the limit setting is acceptable
Upper Sys. has been set
to Min [Max]
The operator has attempted to set the
upper systolic pressure limit higher or
lower than allowed.
Dialog
Message
The machine has set the limit to the highest or lowest value allowed.
Verify that the limit setting is acceptable
Valve 43 Failure
Valve 43 has remained open too long
Alarm
Turn machine power off and back on. Just before beginning dialysis,
verify that the dialysate flow can be turned off and back on. Do not
initiate or continue dialysis if this cannot be done.