Chapter 5—Disinfection and Maintenance
130
2008K Hemodialysis Machine Operator’s Manual Rev P
Acid Clean Program
The 2008K hemodialysis machine should undergo an acid cleaning daily when using
bicarbonate concentrates during dialysis. The purpose of the Acid Clean program is to
prevent the buildup of bicarbonates inside the machine that can have a detrimental effect on
the machine’s performance and treatment efficacy.
The Acid Clean program is not a
method of disinfection.
Acid Cleaning can be accomplished using white distilled vinegar (5% acetic acid) or 2-5%
citric acid.
To run the Acid Clean program:
1.
Attach a sign to the front of the machine that identifies the chemical being used to acid
clean the machine.
2.
Ensure that both dialysate lines are on the shunt and both concentrate connectors are in
their respective ports.
3.
From the “Select Program” screen, touch
Acid Clean
.
4.
Message “Wait: Rinsing Line” appears.
5.
Attach the acid and bicarbonate connectors to a jug (s) containing an acid cleaner when
prompted.
6.
Press
CONFIRM
to start the Acid Clean program. The “Acid Clean” screen appears in
the display (see Figure 59 – The Acid Clean Screen). The progress of the acid cleaning is
indicated by the horizontal bar.
7.
Return connectors to their ports when prompted.
8.
Press
CONFIRM
to exit.
Figure 59 – The Acid Clean Screen