Chapter 6—Alarms and Troubleshooting
149
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
Base Na+ greater than
max. value
Entered Base Na+ higher is than
allowed.
Dialog
Message
The Base Na+ will be set to the highest allowed Na+ level. Press
CONFIRM
to clear message and accept the maximum allowed value.
Verify that the value is acceptable or enter new value.
Base Na+ has been set
to min.
The operator has attempted to set a
Base Na+ lower than allowed.
Dialog
Message
The Base Na+ will be set to the lowest allowed Na+ level. Confirm if
value is acceptable or enter new value.
Bic Pump Alarm
This is a pump failure warning.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, return the blood to the patient if alarm occurs during
treatment. Take the machine out of service.
Bic Pump Always EOS
This is a pump failure warning.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, return the blood to the patient if alarm occurs during
treatment. Take the machine out of service.
Bic Pump No EOS
This is a pump failure warning.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, return the blood to the patient if alarm occurs during
treatment.Take the machine out of service.
Bic Conn Out of Port
The blue bicarbonate wand/connector
is out of its port.
Warning
Connect blue (bicarbonate) wand/connectors into the blue rinse port.
Verify the concentrate selection.
Bicarbonate greater than
max. value
The operator has attempted to set a
Bicarbonate level higher than allowed.
Dialog
Message
The Bicarbonate will be set to the highest allowed bicarbonate level.
Press
CONFIRM
to clear message and accept the maximum allowed
value. Verify that the value is acceptable or enter new value.
Bicarbonate has been set
to min.
The operator has attempted to set a
Bicarbonate level lower than allowed.
Dialog
Message
The Bicarbonate will be set to the lowest allowed bicarbonate level.
Confirm if value is acceptable or enter new value.
Blood flow unstable
When attempting to start an OLC test,
certain conditions are necessary,
including stable blood flow rate.
Advisory
Wait a minute or so and start the OLC test again.
Blood Leak not Calib
The blood leak detector is not in
calibration.
Alarm
Return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.