Chapter 4—Monitoring the Treatment
102
2008K Hemodialysis Machine Operator’s Manual Rev P
The following is information about each of the information lines in the treatment summary
display.
Table 24 – The Treatment Summary Information
Display
Description
Pt ID
This is the patient ID. Currently this is only used by the FDS08
system.
Start Time
This is the clock time when the
Tx Clock
button is touched (24 hour
clock).
End Time
If the treatment is still underway, this is the projected time for the end
of treatment, based on the current time and RTD. Otherwise it is the
clock time when the treatment actually ended, based on the Tx clock
(24 hour clock).
Actual Tx
time
This is the total treatment time, even if the treatment continued after
RTD counted down to zero (minutes).
RTD
The current Remaining Time of Dialysis (minutes).
Hep. Infused
This is the amount of heparin infused to the patient at this point in time
(ml).
Begin BP
Displays the first Diastolic, Systolic (mmHg) and pulse reading
(beats/min). If any of the readings is out of the alarm range, the entire
line is shown in red.
End BP
Displays the last Diastolic, Systolic and pulse reading. If any of the
readings is out of the alarm range, the entire line is shown in red.
Diastolic
Range
Displays the highest and lowest Diastolic pressure reading during the
treatment. If any of the readings is out of the alarm range, the entire
line is shown in red.
Systolic
Range
Displays the highest and lowest Systolic pressure reading during the
treatment. If any of the readings is out of the alarm range, the entire
line is shown in red.
MAP Range
Displays the highest and lowest Mean Arterial Pressure (MAP) during
the treatment. If any of the readings is out of the alarm range, the
entire line is shown in red (mmHg).
Pulse Range
Displays the highest and lowest pulse rate during the treatment. If any
of the readings is out of the alarm range, the entire line is shown in
red.
UF Goal
This is the UF goal selected for the treatment (ml).
UF Removed This is the UF removed at this point in the treatment (ml).
Avg UF Rate
This is the average UF rate at this point in the treatment (ml).
UF Time
This is the UF time selected for the treatment (min).
UF Profile
This is the UF profile # selected for the treatment