Chapter 1—Overview
28
2008K Hemodialysis Machine Operator’s Manual Rev P
Blood Circuit Section
Figure 7 – Blood Circuit Section
The Blood Circuit section contains keys and warning lights that are directly related to the
transmission and monitoring of the patient’s blood. These keys are outlined in red on the
control panel.
Table 3 – Blood Circuit Section Keys
Press …
To …
Keep the blood pump running for three minutes when a blood-leak
alarm is present. The yellow Override light will illuminate.
OR
If a blood leak alarm is not present, pressing and holding the
Override
key for one second will spread the arterial and venous
alarm limits 300 mm Hg and the TMP alarm limits are spread fully
open for 30 seconds. The Override light will not illuminate.
Warning!
During an override, the blood leak detector is
inactive. Monitor the treatment.