Machine Specifications
2008K Hemodialysis Machine Operator’s Manual Rev P
219
Proportional Mixing System
Acid
Volumetric, selectable:
1:34
1:44
1:35.83
1:35.1
Note: Citrasate® is for use with 1:44 concentrates only.
Acetate
1:34
Adjustment Range
130 to 155 mEq/L Na
+
Bicarbonate
Volumetric, selected with associated acid ratio
1:27.46
1:19.13
1:25.16
1:27.6
Adjustment Range
20 to 40 mEq/L Bicarbonate (post-reaction, after mixing with
the acid and purified water).
Monitoring Conductivity
Average Accuracy: ± 1.5%
Method: Temperature compensated electronic conductivity
meter with adjustable alarm limits.
Temperature compensated conductivity display with
automatically set alarm windows ±0.5 mS/cm around
calculated conductivity. User can adjust an additional ±0.5
mS/cm within this range. Conductivity is based on the
concentrates’ compositional data entered in the Dialysate
screen at the standard temperature of 25
C.
Range of Display
10.0 to 17.0 mS/cm. At 25 °C. Alarm limits will not go below
12.5 or above 16.0 mS/cm.
Dialysate Heating
Nominal Value of Temperature
35 to 39 °C Accuracy: ± 0.3 °C (measuring accuracy under
calibration conditions for a dialysate flow of 500 ml/min)
(selectable in 0.1 °C steps)
Temperature Display
Range 35 to 39 °C with alarm limit window automatically
adjusted to 2 °C above and below set point. Alarm window
will not go below 30° or above 41 °C. Heater 1.3 kW,
electronically controlled.
Heat Disinfection
Temperature
83
8 °C at NTC 3
Flow Rate
600 ml/min.
Pre-rinse either 7 min @ 600 ml/min or 20 min @ 300 ml/min
(user selectable). 10 min @ 600 ml/min for DIASAFE PLUS
equipped machines.
Time
Between 10 and 60 minutes (internally selectable)
Auto Heat Disinfect Pre-rinse Time
Between 15 and 30 minutes (user selectable) @ 600 ml/min
(standard) or 350 ml/min (extended pre-rinse). Note: Heater
is off during pre-rinse.
Auto Heat Disinfect Pressure
25 psi < pressure < 90 psi
Note: Silicon inlet/drain tubing set #M38512 must be used
with this option