Chapter 6—Alarms and Troubleshooting
163
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
Fail * Neg Flow On
Negative flow on pressure holding
test failed
Test
Message
Rerun test. If failure repeats, remove machine from service and alert
a qualified service technician.
Fail * Neg Stabilize
Negative flow stabilize test failed.
Test
Message
Rerun test. If failure repeats, remove machine from service and alert
a qualified service technician.
Fail * Optical Detect
Optical Detector test has failed.
Test
Message
Rerun test, if failure repeats, remove machine from service and alert
a qualified service technician.
Fail * Pos Flow Off
Positive flow off pressure holding test
failed
Test
Message
Rerun test. If failure repeats, remove machine from service and alert
a qualified service technician.
Fail * Pos Stabilize
Positive flow stabilize test failed.
Test
Message
Rerun test. If failure repeats, remove machine from service and alert
a qualified service technician.
Fail * (Remove Air)
Remove air test failed.
Test
Message
Rerun test. If failure repeats, remove machine from service and alert
a qualified service technician.
Failed Sending Data to
Actuator Board
Functional to Actuator board
communication problem during
startup.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Fail * Temp High Soft
Fail * Temp Low Soft
Temperature test has failed.
Test
Message
Verify stable temp. Rerun test. If failure repeats, remove machine
from service and alert a qualified service technician to recalibrate.
Fail * Temp High Hard
Fail * Temp Low Hard
Temperature test has failed.
Test
Message
Verify stable temp. Rerun test. If failure repeats, remove machine
from service and alert a qualified service technician to recalibrate.
Fail * TMP High Soft
Fail * TMP Low Soft
Fail * TMP High Hard
Fail * TMP Low Hard
Transmembrane Pressure (TMP) test
has failed.
Test
Message
Rerun test. If failure repeats, remove machine from service and alert
a qualified service technician.