Appendix C—The 2008K@home Hemodialysis Machine
230
2008K Hemodialysis Machine Operator’s Manual Rev P
Figure 75 – 2008K and 2008K@home “Home” Screens
Table 36 – Software Screen Differences
Item
2008K
2008K@home
Setting Treatment
Parameters
Set on various screens,
organized by screen-
buttons along bottom of
display screen
Treatment parameters are grouped on
the “Rx Parameter” screen (accessed
from the “Select Program” screen) and
programmed all at once outside of
Dialysis Mode.
The UF goal is calculated based on the
UF Time (called ‘Tx Time’) and patient
weight entered in the guided “Tx Set up”
screens.
Testing the
Machine
All tests are performed on
the “Test & Options”
screen
Tests are performed during guided
setup; during treatment, tests are
available on the “Help” screen
Setting Dialysate
Parameters
The “Dialysate” screen is
accessed by touching the
Dialysate
screen-button
along the bottom of the
Touchscreen
Concentrates are selected in the “Rx
Parameters” screen.
Concentrates are confirmed in the
guided set up process. Touching the
Conductivity
button on the “Home”
screen will display these settings if limits
need to be adjusted.
UF and SVS
Profiles
Available from the
“Home” screen
The ‘In Center’ option must be set in
order to display the Dialysis Mode
SVS
Profile
and
UF Profile
buttons.
To set the ‘In Center’ option: in Service
Mode, on the “Options: Treatment
Options” screen, touch the
In Center Set
up
toggle button to set it to ‘In Center’
then press the
CONFIRM
key.