Chapter 6—Alarms and Troubleshooting
151
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
Blood Pump Button Alarm Key stuck or held in too long
Alarm
See message A.16
Blood Pump Direction
Error
Pump is turning in the wrong direction
Alarm
See message A.13
Blood Pump Calib Alarm
Pressure was adjusted too much in
calibration mode
Alarm
See message A.26
Blood Pump EEPROM
Err
EEPROM error
Alarm
See message E.05
Blood Pump EPROM
Error
EPROM CRC error
Alarm
See message E.01
Blood Pump Erasing
Error
Error erasing Flash ROM while in
Service Mode
Alarm
See message E.98
Blood Pump Failure
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm does not clear, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified technician.
Blood Pump Flash Error
Error copying data into Flash ROM
while in Service Mode
Alarm
See message E.97
Blood Pump RAM Error
RAM check error
Alarm
See message E.03
Blood Pump Rate Alarm
Pump is not reaching speed at
maximum voltage
Alarm
See message A.11
Blood Pump ROM Error
Flash ROM CRC error
Alarm
See message E.02
Blood Pump Stop Alarm
Pump rotor turning when it should not
be
Alarm
See message A.29