Chapter 6—Alarms and Troubleshooting
Acid Pump Alarm
This is a pump failure warning.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn power off and back on. If warning is not cleared,
return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.
Acid Pump Always EOS
This is a pump failure warning.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn power off and back on. If warning is not cleared,
return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.
Acid Pump No EOS
This is a pump failure warning.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn power off and back on. If warning is not cleared,
return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.
Act Blood Pump Failed
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Act Board CRC Error
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Act BYP Valve Fail 1
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Act BYP Valve Fail 2
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Actuator BD no Echo
Functional to Actuator board
communication problem
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Adjusting TMP
The operator has chosen to relieve
the TMP after a TMP alarm
Advisory
No action necessary.