2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Flow Recirc Error 2
Dialysate flow problem.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Set Dialysate Flow in the “Home” screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min
50 ml/min.
4) Turn the power off and on. If warning is not cleared, return the
blood to the patient if alarm occurs during treatment.
5) Take the machine out of service and alert a qualified service
technician.
Front Panel No Comm
The processor is unable to
communicate with the front panel
Opening
Screen
Message
Turn machine power off and back on. If failure repeats, return the
blood to the patient if alarm occurs during treatment. Take the
machine out of service and refer to a qualified service technician.
Heparin Dwell Complete
The five minute timer for a manual
heparin bolus has elapsed.
Advisory
Press the
RESET
key to clear the message. The Status Light will
stop flashing.
Greater than max. value
Entered parameter is larger than
allowed
Dialog
Message
Verify that the maximum value is acceptable. Press
CONFIRM
to
clear message and accept the maximum allowed value.
Heat Relay Test Fail
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Heparin Pump Alarm
The Heparin pump is encountering
resistance.
Alarm
1) Check the heparin line for clamps or kinks and correct.
2) Check the heparin syringe for adequate amount of heparin and
correct.
3) Ensure the correct type of syringe is loaded and locked in place
properly.
4) Press
RESET
to clear the alarm and restart the heparin pump.
5) If the alarm will not reset or continues to alarm intermittently,
return the blood to the patient if alarm occurs during treatment.
6) Take the heparin pump out of service and alert a qualified service
technician.
Note
: An alarm will sound
when the heparin pump has
reached the end of its
stroke during normal
operation.