150
2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Blood Leak?
The blood leak detector has detected
the presence of blood or air in the
dialysate.
Blood
Alarm
1) Press
RESET
to reset the alarm. If allowed by facility policy,
press
Override
to continue dialysis if the machine cannot be
reset.
2) Check dialysate fluid for presence of blood with a blood leak test
strip.
If test is negative
, recheck with a new blood leak test strip. If
negative after three checks, follow steps below:
Press
Override
to run the blood pump for up to 3 minutes while
troubleshooting the alarm.
Check the dialyzer supply and return lines for air leaks, especially at
the dialyzer connectors and the filter screen in the dialyzer return line.
Press
RESET
to reset alarm.
If unable to reset the alarm, return the patient’s blood according to
procedure below (test positive) and alert a qualified service
technician.
Note
: Air or disinfectants
containing peracetic acid
may cause a false alarm.
Warning
! During an
override, the machine’s
blood leak detector is
inactive. You must monitor
the treatment.
If test is positive
, proceed according to facility blood leak policy. If
facility policy is to return patient’s blood, follow the steps below.
1) Press
RESET
to reset all other blood flow alarms.
2) Press
Override
to enable the blood pump to run and return
patient’s blood per unit protocol.
Note
: Override will activate the blood pump for about
three minutes while a blood leak alarm exists. Press
Override
again if more time is needed to return the
patient’s blood.
Blood Pump +5 V Error
+ 5 volts is outside the allowable
range
Alarm
See message E.10
Blood Pump +12 V Error
+ 12 volts is outside the allowable
range
Alarm
See message E.07
Blood Pump -12 V Error
- 12 volts is outside the allowable
range
Alarm
See message E.09
Blood Pump +24 V Error
+ 24 volts is outside the allowable
range
Alarm
See message E.08