Appendix B
204
2008K Hemodialysis Machine Operator’s Manual Rev P
Adding New Concentrates or Changing the Type
New concentrates can be either selected from a pre-programmed list or by entering the
concentrate manually. This process is performed in Service Mode.
Step 1
Power the machine On into Service Mode by pressing the
CONFIRM
key when prompted
during the power up sequence.
Step 2
Touch the
Options
screen-button and then the
Enter Conc.
screen-button.
Step 3
Verify that the correct family of concentrates is selected. If it is not, select
Change Type
and
select the correct family using the
or
(up/down) keys on the data entry keypad.
Figure 67 – Selecting the Concentrate Type
Touch the
Change Type
button and use
the data entry
keypad to
change the
concentrate type
of dilution to the
desired family