176
2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Recirculating Done
A prompt to the operator that the
recirculation process is done.
Advisory
Press
RESET
to clear the advisory message.
Recirculating Stopped
Recirculation has been stopped
because blood is sensed or the
dialyzer supply and return lines are on
shunt.
Warning
Inspect the configuration of the dialyzer supply and return lines and
extracorporeal blood circuit. Correct any irregularities. If the message
is not cleared, take the machine out of service and alert a qualified
service technician.
Release CONFIRM to
stop
This message is a prompt for the
operator to release the
CONFIRM
key
to stop priming the heparin line.
Dialog
Message
Release the
CONFIRM
key.
Rel. Blood Volume Low
The BVM module has reported a
relative blood volume below the lower
limit.
Alarm
Press
RESET
to clear the message. Evaluate the fluid status of the
patient.
RESET to adjust TMP
The TMP exceeded the hard alarm
limits. The operator is given the
option to relieve the pressure to bring
the TMP within limits.
Warning
Press the
RESET
key to reset the TMP alarm limits. Press and hold
the
Override
key to re-center the limits.
Warning!
Adjusting the
TMP repeatedly will
decrease the UF removed
from the patient.
Warning
! Rising TMP may indicate a leak in the balancing
system and should be investigated.
Reset Treatment?
CONFIRM or Escape
The
New TX
key has been pressed
Advisory
To reset treatment parameters for a new treatment, press the
CONFIRM
key. To cancel, press the
Escape
key.
Resetting, Try Again
Blood pressure module resetting
Warning
Wait until blood pressure module completes resetting and retry blood
pressure reading.
Retry > Press = XXX
The cuff pressure is too low to
measure the blood pressure. The cuff
pressure is XXX mmHg.
Dialog
Message
No action necessary
Reverse bloodlines
This message is a prompt for the
operator reverse the bloodlines for the
Access Flow test
Warning
To proceed with the Access Flow test, reverse the bloodlines and
press
CONFIRM
.
Rinse Cond High
The Reverse Osmosis (RO) water
inlet conductivity is too high.
Opening
Screen
Message
Press the
RESET
key to clear the message. Perform a Rinse cycle. If
alarm is not cleared, take the machine out of service and alert a
qualified service technician.