Chapter 3—Setting Treatment Parameters
82
2008K Hemodialysis Machine Operator’s Manual Rev P
To set an SVS profile:
Figure 39 – Entering an SVS Program
The SVS timer is activated when the
Tx Clock
button is initially selected and confirmed to
start treatment. The green light located above the
SVS on/off
key will illuminate. The SVS
profile parameters can only be changed if the SVS is turned off using the
SVS on/off
key.
1) From the “Home”
screen, touch the
SVS
Profile
button. The SVS
Profile window opens.
2) Select the prescribed profile by touching
the corresponding button from among the
four options located in a row along the top
of the screen.
5) Touch the
SVS Profile
button and, using the data
entry keypad, enter the
running time for the SVS
profile.
3) Touch
Start Na+
,
and enter the
starting maximum
Na value using the
data entry keypad
on the control panel.
4) Touch
Base Na+
, and,
using the data entry keypad,
enter the final Na
concentration value to be
reached at the end of the SVS
profile.
6) Press
CONFIRM
to save
the SVS settings.