Chapter 5—Disinfection and Maintenance
2008K Hemodialysis Machine Operator’s Manual Rev P
133
Figure 61 – The Chemical Rinse Screen
5.
Connect the red connector to a jug containing the chemical disinfectant and press
CONFIRM
when prompted.
6.
Water pre-rinse will start. Remaining Pre-rinse Time meter box will count down.
7.
When Remaining Pre-rinse Time meter box reads 0:00, chemical rinse will start after a
delay. Remaining Time meter box will count down.
Note
: Visually confirm that disinfectant has been pulled into the machine.
8.
When the Remaining Time meter box reads: 0:00, remove the acid connector from the
disinfectant jug and insert it into the acid rinse port when prompted. Post rinse will start
and Remaining Post Rinse Time meter box counts down.
9.
Press
CONFIRM
to exit.
Warning!
Test for residual disinfectant prior to starting treatment following a chemical
disinfection.
Note
: The machine will automatically perform a Diasafe test after the Chemical Rinse
program completes (Functional board software version 3.36 or later).