168
2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Hold CONFIRM to Prime
In heparin screen after selecting
prime must press and hold confirm to
prime heparin line.
Dialog
Message
Press and hold
CONFIRM
to prime the heparin line.
I2C Read Time Out
I2C Bus Read Error
I2C Bus Read Too Long
I2C Byte Write Error
Functional to I2C EEPROM
communication problem.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
In bypass for 8 min
Press
RESET
to clear the message
Advisory
The machine was in bypass for about eight minutes. This may
extend the time necessary to complete the treatment or rinsing of
germicide. Press
RESET
to clear the message.
Warning!
If rinsing germicide from the machine or
dialyzer when this occurs, additional time will be
necessary to fully rinse the germicide from the dialyzer.
Always check for residual germicide using the
appropriate approved residual test method.
INTERRUPT RINSE?
Escape or CONFIRM
Press
CONFIRM
to interrupt rinse.
Dialog
Message
Press
CONFIRM
to accept or press escape to cancel.
Interrupted
The selected Rinse program has
been interrupted.
Warning
Re-insert the dialysate connectors into the proper rinse ports. To
continue the Rinse or other program, press
CONFIRM
, then reselect
the desired program.
Invalid Data Entry for
[item]
Entry value for [item] is out of range
Dialog
Message
Set appropriate value for [item]
Invalid UF Rate
Entry value for goal is out of range.
Dialog
Message
Readjust rate
Invalid UF Time
Entry value for goal is out of range.
Dialog
Message
Readjust time