172
2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Must Run Test First
The Forced Test is required before
proceeding with UF or SVS.
Warning
To proceed with the selected operation, run the Pressure and Alarm
tests. For more information, see Chapter 2 “Testing the 2008K
hemodialysis machine.”
Need Blood Sensed
An action has been initiated that
requires that blood is sensed.
Dialog
Message
Verify venous bloodline is in the Optical Detector.
Neg. Access Flow value
This message is an advisory
message that the Access Flow test
result was a negative value. A
positive value is expected.
Warning
This can occur if the bloodlines were initially connected in the
reversed position. Press
CONFIRM
to clear the message. Check that
the bloodlines are properly connected and repeat the Access Flow
test.
Negative error AF value
This message is an advisory
message that the Access Flow test
result was an erroneous value
Warning
Press
CONFIRM
to clear the message. Check that the bloodlines are
properly connected and repeat the Access Flow test. Press
CONFIRM
to clear the message.
New Art Limits chosen
This message advises the operator
that a new set of arterial limits has
been set.
Advisory
Advisory only. No action is required.
New features loaded,
Power Off, Replace
EEPROM
Advisory message when uploading
hardware key option
Advisory
Put the original calibration EEPROM in IC 20 and restart the
machine.
New TMP Limits Chosen
This message confirms that a new set
of TMP limits have been set.
Advisory
Advisory only. No action is required.
New Venous Limits?
This message is a prompt for the
operator to set new venous alarm
limits.
Advisory
To set new venous alarm limits, press
RESET
and hold.
New Ven Limits Chosen
New venous alarm limits are set
Advisory
Advisory only. No action is required
No Air Detector Alarm
The
Prime
key has been pressed. A
level detector alarm must exist for this
function to occur
Advisory
If the venous chamber has fluid detected, the prime function will not
occur. Press
RESET
to start the blood pump. If a level detector
alarm occurs, then press the
Prime
key.