Chapter 4—Monitoring the Treatment
2008K Hemodialysis Machine Operator’s Manual Rev P
109
How to run the Access Flow test
Figure 50 – Starting an Access Flow Test
When the
Access Flow
button is turned ON, the machine will offer to do the Access Flow
test following the next OLC test. The
Access Flow
button may be left in the Off position and
turned On later. If desired, touch the
Manual OLC Test
and press
CONFIRM
after turning
the Access Flow ON to begin the test right away. If you display the “Kt/V AF” screen while
doing the test, more detailed instructions are displayed.
Warning!
The Access Flow procedure requires that the bloodline connections to the
access needles be reversed and later returned to their original position. To avoid the
possibility of significant blood loss, be sure that the connections are well secured after
disconnecting and reconnecting the bloodlines.
Warning!
Use aseptic technique when doing this procedure.
Warning!
Return the bloodlines to the original position (red to red and blue to blue) when
the test is completed. Failure to do so will result in lower delivered therapy.
Note
: If the access flow rate is less than or equal to the blood pump rate, the access flow
rate will be calculated and reported as approximately the blood pump rate. In this case, the
access flow rate may be lower than indicated.
Note
: During the second OLC measurement for the Access Flow test, the UF will change
to 70 if running low flux or 300 if running high flux.
1) Press the
Access Flow
button and then
CONFIRM
.
2) Press the
Manual OLC Test
button to start the
test now.