Machine Specifications
222
2008K Hemodialysis Machine Operator’s Manual Rev P
Functional Options
Online Clearance(Optional)
Dialysate Flow rate: 300 – 800 ml/min
# of Tests: 1 – 6 during each treatment
Access Flow (Qa)
(Optional, requires OLC)
Minimum Qa: Will not determine the Qa if less than the
blood pump speed.
Maximum Qa: 2000 ml/min
Sodium Variation System (SVS)
A means for temporarily increasing the sodium
concentration at the beginning of dialysis for patient
comfort.
Sodium Variation Profiles
Three preset profiles (step, linear, and exponential) for
increasing, then decreasing the sodium concentration in
dialysate.
Blood Temperature Monitor (BTM)
(Optional)
A means of temperature control for the patient and for
evaluating adequacy of access flow by measuring
temperature changes in the arterial and venous lines
after temporary excursions in the dialysate temperature.
Blood Volume Monitor (BVM)
(Optional)
A module that measures the relative blood volume
(hematocrit) as a means of determining if the fluid
refilling rate from the body to the blood is insufficient to
support the selected ultrafiltration rate. A fast rate of
decrease or steeper slope in the blood volume trend
graph may signal an upcoming hypotensive event.
Diasafe Plus Filter
A means of filtering the dialysate to reduce
bacteriological and endotoxin exposure
Auto Heat Disinfect
Allows the user to program the 2008K hemodialysis
machine to automatically run a heat disinfection program
according to a schedule.
User Interface
Language
The operating screens may be set to either French
(Canadian), Spanish (Mexico), or English (USA)
Blood Pressure Module
Technique
Measures systolic, diastolic pressures, and heart rate
(pulse rate) using oscillometric method. MAP measured.
Cuff Deflation
Interactive computer controlled. Determination for adults
requires approximately 25-30 seconds depending on
starting point, heart rate and motion artifact.
Cuff Inflation
Typically 5 - 10 seconds from 0-250 mmHg.
Interval Settings
Interval times: 5 - 60 minutes in increments of 5 minutes
Clock Time: 5, 10, 15, 20, 30, 60 minutes