2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Conductivity out of range
When attempting to start an OLC test,
certain conditions are necessary,
including the condition that
conductivity for both the inlet and
outlet sensors be in range
Advisory
Wait until the conductivity is stable and start the OLC test again. If
the message repeats, do not use OLC until the conductivity sensors
have been recalibrated.
CONFIRM Concentrate
This message will be displayed if the
user needs to confirm the concentrate
selected for use
Advisory
Press
CONFIRM
or change the concentrate selection and then press
CONFIRM
.
Connector(s) Out Of Port
An action has been initiated that
requires the Acid/Bicarbonate
Connectors to be in their rinse ports
Warning
Insert the concentrate connectors into their proper rinse ports.
Cooling Down
The machine is cooling down from a
heat disinfect.
Advisory
Advisory message only. No action is required.
** Cover is Open **
The dialysate shunt door is open.
Advisory
To proceed with the selected operation, close the shunt door.
Cuff Pressure = XXX
This is displayed during the blood
pressure measurement. The cuff
pressure is XXX mmHg.
Dialog
Message
No action is necessary.
Dial Valve Failure 1
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm does not clear, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Dial Valve Failure 2
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm does not clear, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.