Chapter 5—Disinfection and Maintenance
2008K Hemodialysis Machine Operator’s Manual Rev P
135
8.
Chemical dwell follows after a delay. Remaining Time meter box counts down.
9.
When the Remaining Time meter box reads 0:00, remove the red acid connector from the
disinfectant jug and insert it into the acid rinse port. The machine will automatically run
for about a minute to draw up the disinfectant left in the tubing.
Note
: Visually confirm that disinfectant has been pulled into the machine.
10.
Following the completion of the chemical disinfection cycle, “
Press CONFIRM to exit
”
will display in the Status Line.
11.
Press
CONFIRM
to exit.
Warning!
The mandatory rinse cycle must be completed and a test for residual disinfectant
must be performed prior to the next treatment.