Chapter 6—Alarms and Troubleshooting
165
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
Flow Inlet Error
Float Switch
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Set Dialysate Flow in the “Home” screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min
50 ml/min.
4) Turn the power off and on. If warning is not cleared, return the
blood to the patient if alarm occurs during treatment.
5) Take the machine out of service and alert a qualified service
technician.
Flow is Off
Dialysate flow is off.
Warning
An action has been initiated that requires the dialysate flow to be on.
To proceed with the selected operation, turn the dialysate flow on.
Flow is still On
Dialysate flow is on.
Warning
An action has been initiated that requires the dialysate flow to be off.
To proceed with the selected operation, first turn the dialysate flow
off.
Flow Rate not Set
If the dialysate flow is turned on while
the touch screen flow rate selection is
still “SEQ”, this reminder is displayed
Advisory
Set the dialysate flow rate to the desired value.
Flow Recirc Error 1
Dialysate flow problem.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Set Dialysate Flow in the “Home” screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min
50 ml/min.
4) Turn the power off and on. If warning is not cleared, return the
blood to the patient if alarm occurs during treatment.
5) Take the machine out of service and alert a qualified service
technician.