Chapter 4—Monitoring the Treatment
2008K Hemodialysis Machine Operator’s Manual Rev P
103
Display
Description
Final RBV
This is the last Relative Blood Volume from the BVM, if available (% of
initial value)
Concentrate
This is the concentrate selected for this treatment
SVS profile
This is the Sodium Variation System profile selected for the treatment
Na Base
This is the base Na level for the Sodium Variation System program
(mEq/l)
Na Start
This is the starting Na level for the Sodium Variation System program
(mEq/l). If SVS is not selected, it is the sodium used.
SpKt/V
This is the current Single pool Kt/V (SpKt/V). If the projected Kt/V is
below the acceptable level, the value is shown in red.
eKt/V
This is the current equilibrated Kt/V (eKt/V)
Volume
This is the volume used for the Kt/V calculation (liters)
BVP
This is total blood volume processed (liters).
Kt
This is effective blood volume processed (liters).
Avg Art
Press
This is the average arterial pressure for the treatment (mmHg)
Mean Kecn
The time weighted average of the individual Kecn measurements
Avg Ven
Press
This is the average venous pressure for the treatment (mmHg)
Avg Dial.
Flow
This is the average dialysate flow used for the treatment (ml/min)
Avg blood
Flow
This is the average blood flow used for the treatment (ml/min)
Last Access
Flow
This is the last access flow determination, if available (ml/min)
Avg Recirc.
This is the average of all the recirculation determinations made for this
treatment (%)
Pediatric
This shows whether or not the Pediatric option is set.
High Flux
This shows whether or not the High Flux dialyzer option is set.