Chapter 6—Alarms and Troubleshooting
173
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
No Chemical Intake
During the main program of chemical
rinse or chemical dwell, the machine
cannot get any chemical in the acid
connector.
Alarm
Retry chemical rinse and if problem persist, remove machine from
service and alert a qualified service technician.
No Na
+
Selected
This is a prompt to the operator that a
Start Na
+
value for SVS has not been
set.
Advisory
To proceed with the SVS operation, set a value for Start Na
+
in the
“SVS” subscreen.
No Program Selected
This is a prompt to the operator that a
Profile was not selected.
Advisory
To proceed with an SVS operation, select an SVS Profile from the
“SVS” subscreen.
No SVS Time Selected
This is a prompt to the operator that
the SVS Time has not been set.
Advisory
To proceed with the SVS operation, set the
SVS Time
button in the
“SVS” subscreen.
No Water
A water inlet valve alarm has
occurred. The machine is not
receiving enough water.
Warning
Inspect the treated water source supplying the machine. Correct as
required. If the alarm does not clear, return the blood to the patient if
alarm occurs during treatment. Take the machine out of service and
alert a qualified service technician.
OLC allowed in X minutes There is a minimum waiting period
necessary between OLC tests
Advisory
Wait the indicated time and start the OLC test again.
OLC steps not calculated
In order to do the OLC test, the
machine must calculate the pump
steps necessary to raise and lower
the conductivity for the test. This
cannot be done until stable
conductivity has been achieved.
Advisory
Wait a couple of minutes after the conductivity is stable and start the
OLC test again.
OLC Test Cancelled!
User has cancelled OLC self test or a
condition occurred during test causing
it to cancel.
Advisory
Advisory only no action required
OLC Test Failed
OLC self test failed.
Warning
Restart machine to rerun the OLC self test.
OLC Test Passed
OLC self test passed
Advisory
Advisory only no action required
Online Clearance Self-
test
Machine is running an OLC self-test
Advisory
Advisory only no action required