Machine Specifications
2008K Hemodialysis Machine Operator’s Manual Rev P
223
Performance Limits
Pediatric
Adult
Cuff Pressure Range
5-220 mmHg
10-325 mmHg
Initial Cuff Inflation
125 mmHg or adjusted by
host
180 mmHg or adjusted by
host
Systolic Determination Range
30-200 mmHg
60-260 mmHg
MAP Determination Range
25-140 mmHg
35-220 mmHg
Diastolic Determination Range
10-180 mmHg
30-200 mmHg
Pulse Rate Determination Range
40-240 BPM
40-180 BPM
Cuff Inflation Rate
5 seconds
5 seconds
Determination Time Normal
Approx. 20 seconds
25-30 seconds
Overpressure Cut Off
220 mmHg
325 mmHg
Transducer Drift
Auto Zeroing
Auto Zeroing
Leakage Rate (Max)
3 mmHg/min in 3 minutes
3 mmHg/min in 3 minutes
Pressure Rate Offset
Auto Zeroing
Auto Zeroing
Alarm Preset Values
Internal alarm values preset to provide alarm limits in
the event individual values are not entered
Pediatric/Neonatal
Adult
Systolic
160/80
200/80
MAP
120/70
120/70
Diastolic
100/50
110/50
Pulse
120/50
120/50
Inflation Pressure
Auto
Auto