Chapter 1—Overview
2008K Hemodialysis Machine Operator’s Manual Rev P
33
Figure 10 – The Blood Pump Module
The following table describes the operational features of the blood pump.
Table 6- Blood Pump Features
FEATURE
PURPOSE
Start/Stop Key
Starts and stops the blood pump. Opening the door will also stop the
blood pump.
Pressure Port
Line from arterial drip chamber is connected to a transducer protector
and attached here to provide arterial blood pressure readings.
Level Adjust Key Pressing the
key (level adjust key on the Blood Pump module) will
raise the level of the fluid in the arterial drip chamber.
Display Window Displays the blood flow rate setting in increments of 5 ml/min during
blood pump operation. When the door is open it displays the pump-
segment diameter in mm.
Up/Down Keys
Increases the speed of the pump when Up arrow (
) is depressed,
decreases the pump speed when Down arrow (
) is depressed. When
door is open, simultaneously press the
and
keys and then press
the
or
key to select the pump segment diameter.
Tubing Retainer
A spring-loaded device that secures the pump segment in place.
Display
Window
Blood Pump
Rotor
Start/Stop Key
Arterial
Pressure Port
Level
Adjust Key
Up/Down
Arrow Keys
Tubing
Retainer
Clamp-Panel
Left and
Right Yokes