Chapter 6—Alarms and Troubleshooting
177
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
RTD = Zero
The RTD (Remaining Time on
Dialysis) clock has counted down to
zero.
Warning
Reset the alarm. This message has alerted the operator that the
preset time on dialysis has elapsed (RTD = 0:00). If prescribed
treatment time has not been completed, the operator must take
further action to comply with the prescribed treatment.
Running Diasafe Test
This message is advising the operator
of the status of the Diasafe test
Advisory
Advisory only. No action is required.
Running Online PHT
This message is displayed when the
online Pressure Holding Test is in
progress.
Advisory
Advisory only. No action is required.
Run Access Flow?
This message is a prompt for the
operator begin the Access Flow test.
Warning
To proceed now, press
CONFIRM
. To delay, press
Escape
. To
cancel, go to the “Kt/V AF” screen and toggle the Access Flow check
box to Off.
Select Concentrate
This message is a prompt for the
operator to select a concentrate.
Advisory
To select a concentrate from the menu, use the
or
(
Up/Down)
keys on the control panel to highlight the desired concentrate, and
press
CONFIRM
. For more information, see Chapter 3, “Setting an
Acid/Bicarbonate Type.”
Select new goal or rate
UF time is out of range.
Dialog
Message
Enter a new UF Goal or reduce UF time.
Select new Goal or Time
UF rate is out of range.
Dialog
Message
Enter a new UF Goal or reduce UF rate.
Select Program
This message is a prompt for the
operator to select a program.
Advisory
To proceed, select the desired program and press
CONFIRM
.
Set Arterial Limits
This is a message to re-center the
arterial limits if necessary
Advisory
Press and hold
Reset
for 1 second to re-center the limits.
Set Blood Flow to 300
This message is a prompt for the
operator to set the blood flow rate in
preparation for the Access Flow test
Warning
To proceed, set
Blood Flow
to 300 and press
CONFIRM
.
Set TMP Limits?
This is a message to re-center the
TMP limits if necessary
Advisory
Rising TMP may indicate a leak in the balancing system and should
be investigated.
Press and hold
Override
for 1 second to re-center the limits.