Chapter 6—Alarms and Troubleshooting
175
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
Press ESCAPE To Stop
[Item]
This message is a prompt for the
operator to press the
Escape
key to
stop loading the heparin syringe or
the Rinse program.
Dialog
Message
To proceed with the selected operation, press
Escape
then press
CONFIRM
.
Pressure Test Failed
The pressure test section (PHT) of
the automated Test Sequence has
failed.
Alarm
Reset the alarm and repeat the test. If the failure message is
repeated on retest, take the machine out of service and alert a
qualified service technician.
Priming
The operator has pressed the
Prime
key and initiated the priming function.
Advisory
Advisory only. No action is required.
Put Connectors in Port
The connectors must be in the
machine ports in order to start a
Rinse program
Advisory
Connect the red (acid/acetate) and/or blue (bicarbonate) connectors
to the appropriate rinse ports.
Put Lines On Shunt
An action has been initiated that
requires the dialyzer supply and
return lines to be on the shunt.
Warning
To proceed with the selected operation, place dialyzer supply and
return lines on the shunt.
Put Red Con in Chemical This is a cleaning/disinfectant
program prompt to the operator.
Advisory
Remove the red connector from the machine and place it into the
wand in the yellow chemical/disinfectant bottle.
RAM Battery Failure
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
RAM Code Corrupted 1
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
RAM Code Corrupted 2
Repeated electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Recirc Interrupted
The Recirculate program has been
interrupted by an alarm condition.
Warning
1) Inspect the blood pump condition. Correct if required.
2) Reset the alarm and turn UF back on, if applicable.
3) If the alarm does not clear, take the machine out of service and
alert a qualified service technician.
Recirculating
Recirculation is in progress.
Advisory
Advisory only. No action is required.