2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
TMP is Low (alarm at or
below 60)
The TMP has exceeded the TMP low
alarm limit value.
Blood
Alarm
1) Ensure that the venous transducer protector is dry and the
monitor line is open. Replace the transducer protector, if
necessary.
2) Check the dialyzer supply and return lines for kinks.
3) Check the filter screen in the dialyzer return line to make sure it is
clean.
4) Press
RESET
key to reset alarm. Press the
Override
key and
hold for one second to select new alarm limits or for adjusting the
TMP.
Note
: Increasing the UF rate can also raise the TMP.
Administer saline as prescribed. Notify a physician if the
UF rate has changed.
Note
: Lowering the venous pressure by reducing the
blood flow rate can also be effective, if using a high-
permeable dialyzer. Notify a physician if the blood flow
rate has changed.
5) If unable to reset the alarm, return the blood to the patient if
alarm occurs during treatment. Take the machine out of service
and alert a qualified service technician.
Tx Clock Paused?
Blood is sensed in optical detector
while Tx clock is paused
Warning
Start Tx clock
UF Goal greater than
max. value
Entry value for goal is out of range.
Dialog
Message
Readjust UF Goal
UF Goal Reached
This message is to alert the operator
that the preset ultrafiltration goal has
been reached.
Warning
Press
RESET
to reset the alarm. The preset UF Goal has been
reached and the UF Rate will drop to the minimum UF Rate. If the
patient’s prescribed UF Goal has not been reached, the operator
must take further action to comply with the prescribed treatment.
UF Is On
An action has been initiated that
requires the UF to be off.
Advisory
To proceed with the selected operation, turn the UF pump off.
UF Profile Error
A UF profile calculation error has
been detected
Alarm
Reset the UF parameters