170
2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Low Temperature
The actual dialysate temperature has
exceeded the low-temperature alarm
limit. Machine is in bypass mode.
Dialysate
Alarm
1) Check that the machine is in Dialysis mode and the dialysate flow
is on.
2) Check that the heater switch on the back panel is in the on (
)
position.
3) Check the water supply to the machine for excessively cold
temperature and correct.
4) Check the Temperature value in the “Home” screen. Re-enter it if
necessary and allow five minutes for the temperature to stabilize.
5) If unable to attain the prescribed temperature, return the blood to
the patient if alarm occurs during treatment.
6) Take the machine out of service and alert a qualified service
technician.
Note
: If the temperature
fluctuates between HIGH
TEMPERATURE and LOW
TEMPERATURE, see
“VARIABLE
TEMPERATURE.”
Lower Dia. has been set
to Min [Max]
The operator has attempted to set the
lower diastolic pressure limit higher or
lower than allowed.
Dialog
Message
The machine has set the limit to the highest or lowest value allowed.
Verify that the limit setting is acceptable
Lower Pulse has been set
to Min [Max]
The operator has attempted to set the
lower pulse rate limit higher or lower
than allowed.
Dialog
Message
The machine has set the limit to the highest or lowest value allowed.
Verify that the limit setting is acceptable
Lower Sys. has been set
to Min [Max]
The operator has attempted to set the
lower systolic pressure limit higher or
lower than allowed.
Dialog
Message
The machine has set the limit to the highest or lowest value allowed.
Verify that the limit setting is acceptable
Max UF rate reached.
Select new Goal or Time
This message informs the operator
that the calculated UF rate is higher
than the internal selection allows.
Dialog
Message
In the “Home” screen, decrease the UF Goal or increase the UF
Time.
Max UF time reached.
Select new Goal or Rate
This message informs the operator
that the calculated UF time is higher
than the maximum allowed.
Dialog
Message
In the “Home” screen, decrease the UF Time