Chapter 3—Setting Treatment Parameters
2008K Hemodialysis Machine Operator’s Manual Rev P
79
To set the 2008K hemodialysis machine for sequential
dialysis:
Figure 37 – Setting Sequential Dialysis
After 60 minutes of pure ultrafiltration, an alarm sounds and the warning message, “
60
MINUTES FLOW OFF
,” appears in the Status Box. The operator has the option of
continuing pure ultrafiltration or starting dialysis. This alarm occurs only once.
To continue pure ultrafiltration,
press the
RESET
key on the left side of the control panel.
This will silence the alarm and terminate the warning message
.
To start the treatment
, touch the
Dialysate Flow
button in the “Home” screen, set it to the
prescribed rate using the data entry keypad, press
CONFIRM
, and start the dialysate flow by
pressing the
Dialysate Flow on/off
key in the upper, right corner of the control panel. When
dialysate flow is on, the yellow light is not illuminated.
3) Touch
Dialysate
Flow
to highlight it.
5) Press
CONFIRM
This
will deactivate the “Flow
Off” warning for five
minutes. The Dialysate
Flow button will display
“SEQ”.
1) Set UF treatment
parameters. Set the UF
Time to include the
combined treatment and
pure UF times.
4) Use the
number or arrow
keys to enter 0.
2) Turn off Dialysate
Flow. Yellow light
will flash when flow
is off.