Machine Specifications
220
2008K Hemodialysis Machine Operator’s Manual Rev P
Chemical Disinfection
Temperature
37 °C (set point applicable)
Flow Rate
620 ml/min
Time
Between 10 and 60 minutes (internally selectable)
Blood Pump
Display of flow rate
8 mm blood line: 20 – 600 ml/min
6.35 mm blood line: 20 – 465 ml/min
4.8 mm blood line: 10 – 274 ml/min
2.6 mm blood line: 6 – 86 ml/min
Accuracy: ± 10% tested at - 200 mmHg
Internal diameter of pump segment
2.6 to 10 mm (0.1” to 0.4”)
Tube length
32 cm minimum (12-5/8”)
Min. pump segment wall thickness
1.26 mm
Durometer
80 shore A nominal
Level adjust
Up only
Power outage use
The pump can be manually operated with a hand crank.
Single Needle System
Two Pump Procedure
With two blood pumps, pressure control system with
alternating blood pumps. Alarm after 15 or 30 seconds
without an alternation of the pumps.
Heparin Pump
Type of Syringe
10 – 12 ml disposable syringe
Administration Rate
0.1 to 9.9 ml/hr
Accuracy: ± 5%
Monitoring
Monitoring end of stroke
Bolus
From 0 to 9.9 ml volume
Monitoring Elements: Blood Circuit
Arterial Pressure Monitor
-300 to +500 mmHg with 3 automatically set time-
delayed alarm window limit values (± 60, ± 80, and ± 100
mmHg of actual pressure (Single Needle
80 mmHg).
Venous Pressure Monitor
-80 to +500 mmHg with 3 fixed window limit values of
±60, ±80, and ± 100 mmHg of actual pressure. There is
also an asymmetric range initially set to ±80 mmHg
which increases the lower limit after 60 seconds (Single
Needle ± 80 mmHg).
Accuracy
± 20 mmHg or ± 10% of indicated reading, whichever is
greater