Chapter 6—Alarms and Troubleshooting
179
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
SN BP Pressure Alarm
Arterial pressure-read back analog
voltage at X348/7 is out of limits
SN pressure-read back analog
voltage is out of limits
Alarm
See message A.22
SN BP RAM Error
RAM check error
Alarm
See message E.03
SN BP Rate Alarm
Pump is not reaching speed at
maximum voltage
Alarm
See message A.11
SN BP Receiving Alarm
Error in received Intel-Hex-line
Alarm
See message A.28
SN BP ROM Error
Flash ROM CRC error
Alarm
See message E.02
SN BP Rotation Error
Pump rotor turning when it should not
be for a second time
Alarm
See message E.23
SN BP Set Rate Alarm
Set speed-read back analog voltage
at X348/14 is out of limits
Alarm
See message A.20
SN BP Stop Alarm
Pump rotor turning when it should not
be
Alarm
See message A.29
SN BP Tach Alarm
Optical tachometer not in range
Alarm
See message A.24
SN BP Task Error
Software task was not completed
correctly
Alarm
See message E.15
SN BP Timer Error
50 ms second time period exceeded
Alarm
See message E.14
SN BP Update Error
Transmit error during Flash update
while in Service Mode
Alarm
See message E.99
SN BP Volt Error
Reference Voltage error
Alarm
See message E.04
SN BP WD Error
Watchdog timeout
Alarm
See message E.06
SN pump in use
The OLC test may not be run when
the Single Needle system is in use.
Advisory
Do not attempt an OLC test when using Single Needle
Standby for Test
This message is displayed before the
start of the Alarms and Pressure test
Advisory
Advisory only. No action is required.
Start Na+ greater than
max. value
Entered Start Na+ parameter is larger
than allowed.
Advisory
The Starting Na+ will be set to the highest allowed Na+ level. Press
CONFIRM
to clear message and accept the maximum allowed value.
Verify that the value is acceptable or enter new value.