Chapter 6—Alarms and Troubleshooting
155
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
BPM: Pulse Amp Unif
The amplitude of the pressure pulses
is inconsistent with an accurate blood
pressure profile.
Blood
Pressure
Alarm
Check pressure cuff for proper fit and alignment. Observe the patient
for physiologic changes. Treat as patient as symptoms warrant.
BPM: Pulse High
The latest pulse reading is above the
Upper Pulse alarm limit.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Pulse Low
The latest pulse reading is below the
Lower Pulse alarm limit.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Pump On > 30 sec
The pump that inflates the cuff has
been running longer than 30 seconds
for an adult patient or longer than 10
seconds for pediatric patient.
Blood
Pressure
Alarm
Cuff is not inflating. Check for loose tubing connections or a leak in
the cuff.
BPM: Systolic High
The systolic blood pressure reading is
above the set Upper Systolic alarm
limit.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Systolic Low
The systolic blood pressure reading is
below the set Lower Systolic alarm
limit.
Blood
Pressure
Alarm
Observe the patient for physiologic changes. Treat patient as
symptoms warrant.
BPM: Weak Pulse
The pulse pressure is too weak to
register an accurate measurement.
Blood
Pressure
Alarm
Check the cuff for proper fit and inflation. Observe the patient for
physiologic changes. Treat patient as symptoms warrant.
BPM: Zero Pressure
No pressure is detected by the blood
pressure module.
Blood
Pressure
Alarm
Check for a loose connection in the inflation system. Correct as
necessary. If no leak found, turn the power off, then back on. If
problem persists, alert a qualified service technician.
BTM test underway
The OLC test may not be started
when a BTM recirculation test is
underway.
Advisory
Wait for the BTM recirculation test to complete before beginning the
OLC test.
BVM Failed
The BVM module has failed
Alarm
Press
RESET
to clear the message. BVM will no longer pass
information to the 2008K monitor. Turn the power off and back on. If
the alarm is not cleared, refer to a qualified service technician.