General Warnings
18
2008K Hemodialysis Machine Operator’s Manual Rev P
Using a Central Venous Catheter
Maintenance
Assembly, installation, adjustment, or repair is to be performed only by persons authorized by
the facility medical director or by Fresenius USA, Inc.
Questions?
For further information regarding the operation, repair, parts, or maintenance of the 2008K
hemodialysis machine, please contact:
Fresenius USA, Inc.
(800) 227-2572
Attention: Service Department
4040 Nelson Avenue
Concord, CA 94520
www.FMCNA.com
Shock Hazard
: Ensure that no conductive electrical devices connected to or near the patient
have leakage currents above the maximum CF applied parts limit of 10 μA DC and 50 μA DC
in a single fault condition. Failure to follow these precautions may result in serious injury or
death.