Machine Specifications
2008K Hemodialysis Machine Operator’s Manual Rev P
221
TMP Monitor
+60 to –520 mmHg with automatically set time delayed
window limit values of
60 (conventional dialysis) and
40 mmHg (high flux dialysis). Compensation for upward
drift.
Level detector
Ultrasonic impulses detect fluid level in the drip chamber.
Optical Sensor
Optical transmission used to detect opaque or non-
opaque presence in the blood tubing.
Clamp
Closes with any blood alarm
Level Adjust
Allows the level in the drip chamber to rise to maintain
the desired fluid level in the drip chamber
Blood Leak Detector
Two color light source transmitter / sensor with a
resolution of:
minor > 0.35 ml/min of blood (hematocrit = 25%)
alarm >0.45 ml/min of blood (hematocrit = 25%)
Ultrafiltration Control
UF Pump Volume Accuracy
± 1% (for P
di
> -500 mbar)
where P
di
= dialysate pressure on the inlet side of the dialyzer
Fluid Removal Rate from Patient
0 – 4000 ml/hr
Dialysate flow rate at 100 ml/min: Accuracy (on total
volume removed): ±(1% UF rate + 18 ml/hr)
Dialysate flow rate at 500 ml/min: Accuracy (on total
volume removed): ±(1% UF rate + 30 ml/hr)
Dialysate flow rate at 800 ml/min: Accuracy (on total
volume removed): ±(1% UF rate + 48 ml/hr)
Adjustment Range of UF Rate
Volumetric Control, 0-4000 ml/hr
1000, 2000, 3000, and 4000 ml/hr internally maximum
rate. Adjusted in 10 ml increments
UF Time
Digital Display (0-9:59 hrs.) Selectable in increments of 1
min.
UF Goal
Digital Display (0-9,990 ml). Selectable in increments of
10 ml.
UF Profiles
Eight UF profiles are available for the removal of fluid
from the patient. Four are preset and four may be
defined by the user.
Remaining Time of Dialysis (RTD)
0-9:59 hours auto transfer from UF time, counting down
in 1-minute increments. Can adjust manually.
UF Removed Display
Digital display max 9,999 ml counting in 1 ml increments.
Additional Monitoring
Alarm in case of power failure.
Alarm in case of water shortage.