Chapter 6—Alarms and Troubleshooting
145
2008K Hemodialysis Machine Operator’s Manual Rev P
Message
Purpose of Message
Type
Action Required
A.25 (Arterial or SN
Blood Pump Message)
Pressure increase when the Level Up
key is pressed
Alarm
Press the
RESET
key to clear. Possibility that the level adjust pump
is connected backward so that the level is lowered instead of raised.
Verify that the level in the arterial chamber rises when the adjust key
is pressed. If it does not, return the blood to the patient if alarm
occurs during a treatment. Take the machine out of service and alert
a qualified service technician.
A.26 (Arterial Blood
Pump Message)
Pressure was adjusted too much in
calibration mode
Alarm
Press the
RESET
key if this message occurs in dialysis mode. If this
alarm occurs during a treatment, return the blood to the patient. Take
the blood pump module out of service and alert a qualified service
technician.
A.27 (Arterial Blood
Pump Message)
Time out when receiving Intel-Hex-
line or overflowed received buffer
Alarm
Press the
RESET
key if this message occurs in dialysis mode. If this
alarm occurs during a treatment, return the blood to the patient. Take
the blood pump module out of service and alert a qualified service
technician.
A.28 (Arterial Blood
Pump Message)
Error in received Intel-Hex-line
Alarm
Press the
RESET
key if this message occurs in dialysis mode. If this
alarm occurs during a treatment, return the blood to the patient. Take
the blood pump module out of service and alert a qualified service
technician.
A.29 (Arterial Blood
Pump Message)
Pump rotor turning when it should not
be
Alarm
Press the
RESET
key to clear. If problem persists, manually return
the blood to the patient (see page 118 for instructions). Take blood
pump module out of service and alert a qualified service technician.
Access Flow Complete
This message is an advisory
message that the Access Flow test is
complete
Warning
Press
CONFIRM
to clear the message
Access Flow Running
This message is an advisory
message that the Access Flow test
process is continuing.
Advisory
No action is necessary
Access Flow Test
Scheduled
This message is an advisory
message that the Access Flow test
process is continuing.
Advisory
No action is necessary
Acetate Selected!
Acetate concentrate has been
selected and the blue bicarbonate
wand/connector is out of its port.
Warning
Connect blue (bicarbonate) wand/connector into the blue rinse port.
Be sure the concentrate selection is correct.