Chapter 4—Monitoring the Treatment
112
2008K Hemodialysis Machine Operator’s Manual Rev P
Blood Temperature Monitor / Blood Volume Monitor Screen
Figure 52 – BTM
and BVM
Monitoring Screen
The Blood Temperature Module (BTM) is an optional and separate device with its own
Operator’s Manual. For a complete understanding of the functions of the BTM, please refer
to P/N 470164. The BTM functions utilize the keys on the module itself for operation. The
Touch Screen is used only for displaying the results and operations of the BTM; none of the
parameters are entered outside the BTM module.
The Blood Volume Module (BVM) is an optional and separate device with its own
Operator’s Manual. For a complete understanding of the functions of the BVM, please refer
to P/N 490041. The BVM functions utilize the keys on the module itself for operation. In
addition, the Touch Screen is used to display a graphical representation of the blood volume
over time and to select the alert level where an alarm will occur.
BTM function
The BTM has two primary functions – to regulate the patient’s temperature (energy) and to
use temporary changes in dialysate temperature to determine the extent of recirculation at the
blood access site.