2008K Hemodialysis Machine Operator’s Manual Rev P
Chapter 6—Alarms and Troubleshooting
Online Clearance Test
Machine is running an OLC
measurement
Advisory
Advisory only no action required
Online PHT Failed
The online Pressure Holding Test has
failed.
Warning
Reset the alarm. Check the machine for liquid leaks. If the failure
message is repeated on the next test (12 minutes between tests),
return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.
Online PHT Too Long
Electronic self-test failure.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Patient Alarm
External alarm
Alarm
Clear external alarm. If problem persists, return the blood to the
patient if alarm occurs during treatment. Take the machine out of
service and alert a qualified service technician.
PHT is running
The PHT test must complete before
the OLC test is allowed to run.
Advisory
Wait 15 seconds and start the OLC test again.
Plug in Venous Pump
Single Needle option was initiated but
the single-needle, blood pump is not
plugged into the machine port.
Warning
To proceed with the selected operation, install the single-needle
blood pump into the machine.
Power Failure Recovery
The machine is powering up after a
power failure. Parameters have been
recovered
Opening
Screen
Message
Verify that all treatment settings are correct before resuming dialysis.
Press CONFIRM to exit
This message is a prompt for the
operator to press the
CONFIRM
key
to exit the Rinse program.
Advisory
To proceed with the selected operation, press
CONFIRM
.
Press CONFIRM to Load
This message is a prompt for the
operator to press the
CONFIRM
key
to load the heparin syringe.
Dialog
Message
To proceed with the selected operation, press
CONFIRM
.
Press CONFIRM to Start This message is a prompt for the
operator to press the
CONFIRM
key
to start the program.
Advisory
To proceed with the selected operation, press
CONFIRM
.
Press ESCAPE to cancel
rinse
This message is a prompt for the
operator to press the
Escape
key to
cancel the Rinse program.
Advisory
To proceed with the selected operation, press
Escape
then press
CONFIRM
.