6-60
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
2.
, press [CALIBRATION].
3.
If necessary, press [CLOSED SAMPLER] to display the Closed
Mode Calibration Factors.
4.
If desired, press [PRINT] to obtain a printout of the <CURRENT
WHOLE BLOOD FACTORS> displayed on the screen.
NOTE: These calibration factors are retained in the
until 10 entries have been made. As each
subsequent entry is made, the oldest existing entry is deleted.
5.
Press [AUTO-CALIBRATE] to select the
6.
If necessary, press [CHANGE SAMPLER] to select the Closed
mode.
7.
Press [WHOLE BLOOD] to select Whole Blood Calibration.
NOTE: If necessary, press [CONFIRM SELECTION] to
continue.
Entering the Reference Values
Enter the reference values for the whole blood samples as follows:
1.
Delete any existing values: press [CLEAR REF VALS] followed
by [CONFIRM CLEAR].
2.
For each sample used, type the appropriate information in the
<SPEC ID> and <# OF RUNS> fields. Press the Enter key on the
keyboard after each entry to save it and advance the cursor.
NOTE: The <# OF RUNS> entered indicates the number of
times the sample will be consecutively cycled through the
instrument. Each sample may be run a maximum of 5 times. For
Mode to Mode calibration, each sample should be run once but
samples may be run more than one time if desired. However,
samples should be run the same number of times in each mode.
3.
Type the Open mode value for each parameter in the appropriate
parameter reference values field. Press the Enter key on the
keyboard to save the entry and advance the cursor.
4.
If desired, press [PRINT SUMMARY] to obtain a printout of the
reference values.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......