Chapter 5 Operating Instructions
Closed Mode QC
QC samples may be run on the Sample Loader using the “Q Labels,”
which are bar code labels provided with the Sample Loader. Each label is
designated Qx, where x indicates the file number. If these labels are
placed on the control tubes, the results are automatically transmitted to
the file indicated by the label. QC samples may also be run on the
Sample Loader without these labels.
NOTE: Be sure the Work List is OFF before beginning these
procedures. If necessary, refer to the
chapter for instructions.
QC (with Q Labels) Procedure
1.
Label each control tube with the Q label that indicates the desired
QC file.
2.
Be sure that the word “Ready” on the Analyzer status indicator
panel is illuminated in green and the READY message is displayed
in the Status Box on the
3.
Place the labeled tubes in the Sample Loader end rack and load the
rack in the Sample Loader tray.
4.
If necessary, press [RUN] to display the
5.
If necessary, press [CHANGE SAMPLER] to select the Closed
mode.
6.
Be sure that all 10 racks (5 on each side) and the safety cover are in
place and press the Start key on the Sample Loader.
7.
The Sample Loader reads the Q label and transmits the results to
the appropriate QC file.
8.
When all controls have been run, press [MAIN] followed by [QC
LOGS].
9.
Use the Arrow keys on the keyboard to move the cursor to the
desired file.
10.
Press [VIEW QC LOG] to display the QC log.
11.
Verify that the results are acceptable.
NOTE: Out of control results are displayed in color.
12.
Repeat steps 8–10 for all controls that were run.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......