CELL-DYN
3000 System Operator’s Manual
5-65
9140240E — May 1995
Chapter 5
Operating Instructions
The [RESISTANT RBC] key is used to initiate the Resistant RBC cycle.
This cycle is used to process samples containing RBCs that are resistant
to the lytic action of the Sheath reagent. The sample is held in the WBC
mixing chamber for approximately 15 seconds longer than the normal
mixing time. The extra time enhances the osmotic lysing effect of the
Sheath reagent and reduces interference from the lytic-resistant RBCs.
NOTE: The interference caused by these RBCs frequently
generates WBC and NRBC flags. The Resistant RBC cycle
reduces the number of these flags generated. Normal samples
run in the Resistant RBC Mode will have an increase in the false
positive band flag rate.
NOTE: Specimens with suspected fragile WBC populations
should not be run in the Resistant RBC Mode.
The key is available only when the Open Mode is selected.
Figure 5.33:
Run Screen for Resistant RBC
RESISTANT
RBC
CLEAR
ORIFICE
WORK
LIST
SPECIMEN
TYPE
CUSTOMIZE
REPORT
CHANGE
SAMPLER
TICKET
REPORT
MAIN
RUN
Ready
May 03 1995
Operator ID
Sequence #
08:05
112
740
Next ID___________Auto
ResRBC___________
Sex(M/F):- DOB:--/--/--
Dr__________________
Param: 1 Limits: 1
WBC:
K/uL
NEU:
%N
LYM:
%L
MONO:
%M
EOS:
%E
BASO:
%B
RBC:
M/uL
HGB:
g/dL
HCT:
%
MCV:
fL
MCH:
pg
MCHC:
g/dL
RDW:
%
PLT:
K/uL
MPV:
fL
D
E
P
O
L
S
I
Z
E
COMPLEXITY
ORTHOGONAL
RBC
PLT
Open Sampler
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......