Sample Loader
The Sample Loader bar coded rack ID label serves as the identification of
tube position number 1. Bar coded position ID labels, numbers 2 through
10, used for identification of the remaining tube positions, are encoded as
letters of the alphabet. The position ID label is read by the bar code
reader during each rack movement. Failure to read the appropriate
position ID label will result in an immediate fault and the Sample Loader
will halt. The message
is displayed on the bulletin line. Refer to
NOTE: Use of specimen ID bar code labels containing only a
single letter or double letters (A to ZZ) will generate an Auto-
Sampler Alarm Condition <u>; See DIAG fault.
The Sample Loader provides positive specimen identification with a
laser bar code reader at Station 2. As each tube reaches Station 2, the
tube rotates and the bar code reader turns on to read the bar code label on
the tube. If no bar code label is present or the label is unreadable, the
sample is identified by the rack number and tube position, in the format
R(n)T(n), on the Data Station RUN Screen and in the Data Log.
The use of bar code labels is recommended for positive sample
identification. Operation of the Sample Loader with non-bar-coded
samples is allowed but careful attention to rack number and tube position
of each sample is required for correct identification.
CAUTION: Liquid spills in the rack drive mechanism are a
potential reason for failure of the rack to advance. Liquid spills
that flow in the Sample Loader Control Panel could cause
operational failure. If a spill occurs, turn the Sample Loader off
immediately and notify the Customer Support Center for further
assistance.
CAUTION: Do NOT remove any non-bar-coded samples from
the racks until the specimen ID numbers are matched to the rack
number and tube position and entered on the record.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......