6-88
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
* Current factor as printed in this manual calibration procedure.
** If factor exceeds limits, do not calibrate. Check all calculations and call the Customer Support Center for assistance at 1 (800)
CELL DYN.
* Mean after calibration.
** If % Diff exceeds limits, call the Customer Support Center for assistance at 1 (800) CELL DYN.
New Closed Mode Calibration Factors
Open Mode Mean x Current Closed Mode Cal Factor* = New Closed Mode Cal Factor
Closed Mode Mean
Open Mode
Mean
/
Closed
Mode
Mean
x
Current
Closed Mode
Cal Factor*
=
New
Closed
Mode Cal
Factor
Range**
WBC
/
x
=
0.700 - 1.300
RBC
/
x
=
0.800 - 1.200
HGB
/
x
=
0.700 - 1.300
MCV
/
x
=
0.700 - 1.300
PLT
/
x
=
0.700 - 1.300
Mode to Mode Post-Calibration Difference
Closed Mode Mean - Open Mode Mean x 100 = % Diff
Open Mode Mean
Closed
Mode
Mean*
-
Open
Mode
Mean
/
Open Mode
Mean
x 100 =
%Diff
Range**
WBC
-
/
x 100 =
<1
.75%
RBC
-
/
x 100 =
<1
.25%
HGB
-
/
x 100 =
<1
.25%
MCV
-
/
x 100 =
<1
.25%
PLT
-
/
x 100 =
<
3.50%
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......