CELL-DYN
3000 System Operator’s Manual
6-49
9140240E — May 1995
Chapter 6
Calibration
NOTE: Comments may be added to the Calibration Log only
after a calibration factor is changed or reentered.
6.
If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Calibration for Individual Parameters
If only certain individual parameters require calibration, continue with
this section:
1.
Press [QUIT AUTO-CAL] followed by [CONFIRM QUIT] to
exit Auto-Cal.
2.
Press [RETURN] twice to return to the
3.
Press [ENTER FACTOR].
ENTER WHOLE BLOOD FACTOR Screen
4.
Type any mean factors computed by Auto-Cal for those parameters
that require calibration in the appropriate fields. (Refer to the
printout made in
Calibration Factor Calculation
section.) Move the cursor to the desired position, type the factor
and press the Enter key on the keyboard to save the entry and
advance the cursor.
5.
Press [RETURN].
6.
parameters that required calibration in the <COMMENTS> line.
Press the Enter key on the keyboard to save the entry and advance
the cursor.
NOTE: Manually entered factors are followed by the letter E in
parentheses to indicate “Enter Factor”.
7.
If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Completing Whole Blood Open Mode Calibration
1.
Press [RETURN] to return to the
2.
Press [MAIN] to exit the
3.
Confirm the calibration of the Open mode by running commercial
controls as directed in the
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......