6-42
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
Auto-Cal Using Whole Blood Samples
Procedure
The calibration procedure has been divided into subsections that consist
of a series of easy-to-follow steps. Always follow the entire procedure
unless specifically directed to skip to another section.
NOTE: Be sure to complete the
sections before beginning this procedure.
Starting Auto-Cal
1.
, press [CALIBRATION].
2.
If necessary, press [OPEN SAMPLER] to display the Open Mode
Calibration Factors.
3.
If desired, press [PRINT] to obtain a printout of the <CURRENT
WHOLE BLOOD FACTORS> displayed on the screen.
NOTE: These calibration factors are retained in the
until 10 entries have been made. As each
subsequent entry is made, the oldest existing entry is deleted.
4.
Press [AUTO-CALIBRATE] to select the
5.
If necessary, press [CHANGE SAMPLER] to select the Open
mode.
6.
Press [WHOLE BLOOD] to select Whole Blood Auto-Cal.
NOTE: If necessary, press [CONFIRM SELECTION] to
continue.
Entering the Reference Values
Enter the reference values for the whole blood samples as follows:
1.
Delete any existing values: press [CLEAR REF VALS] followed
by [CONFIRM CLEAR].
2.
Type the appropriate information in the <SPEC ID>, <# OF
RUNS> and parameter reference values fields for each whole
blood sample to be used for the calibration. (A minimum of 5
whole blood samples is required. Each sample must be run in
triplicate.) Press the Enter key on the keyboard after each entry to
save it and advance the cursor.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......