CELL-DYN
3000 System Operator’s Manual
10-57
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.2
General Fault Conditions (Continued)
Diluent, Lyse, or Sheath empty/SB and BL
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
Container is empty.
1.
Install a new container of
reagent and then press
[CLEAR FAULT].
NOTE: Do not pour any remaining reagent into the new
container.
2.
Reagent inlet tubing is
2.
Inspect the inlet tubing
crimped or obstructed.
to ensure it is not crimped
and/or remove any
obstruction.
3.
Reagent line is not on
3.
Ensure that the line is
the bottom of the
properly inserted in the
container.
container and the sinker is
on the bottom of the
container.
4.
An incorrect reagent or a
4.
Check the label on the
nonconductive liquid is
reagent container to be sure
connected to the inlet tube.
the correct reagent is
installed. Trace the line to
the inlet connector and
ensure that it is connected
to the correct one. Check
the connection to be sure it
is secure and then press
[CLEAR FAULT].
5.
Circuitry malfunction.
5.
Call for technical assistance.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......