5-78
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
6.
When the cycle is completed, the probe moves down, the word
“Ready” on the Analyzer status indicator panel is illuminated in
green and the results are displayed on the
7.
If automatic report printing has been specified, a report is printed
according to the options selected during set up. If this has not been
specified, a report may be printed by pressing [PRINT REPORT].
8.
Repeat this procedure for subsequent samples.
Closed Mode Analysis
The Closed Sampler Mode on Sample Loader instruments aspirates the
sample from a closed collection tube that has been placed in a Sample
Loader rack and loaded into the Sample Loader. CLOSED SAMPLER
is displayed in the upper right corner of the
when this mode is selected. The bulletin line displays the message
AUTO-SAMPLER READY and the Sample Loader Start key indicator
blinks when the Sample Loader is ready. The probe is retracted when this
mode is selected.
NOTE: The last group of samples in a run should be placed in
the Sample Loader rack known as the end rack, which
automatically signals the Analyzer to stop processing. This rack
has black labels on the top and a black bar on the left edge.
Closed Mode Procedure
1.
Be sure that the word “Ready” on the Analyzer status indicator
panel is illuminated in green and READY is displayed in the Status
Box on the
. The bulletin line should display the message
AUTO-SAMPLER READY.
2.
press [CHANGE SAMPLER]
to select the Closed Mode.
3.
Place the well-mixed samples in the Sample Loader racks in the
order in which they are to be run.
NOTE: The racks and tube positions are identified by the bar
code label on the rack and indicated as R(n)T(n). These numbers
appear as the Sample ID number if bar code labels are not used.
(If the Work List is used, the Sample ID number is taken from it.
Refer to the
section of this chapter for more
information.)
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......